Measuring technology: Fluxmeter

M-Flux 1000

Measurement of total Flux of permanent Magnets and Magnet Systems

M-Flux 3000

3-dimensional dipole measurement of permanent magnets

Specifications

Technical data

Measurement ranges

± 500µVs, ± 5mVs, ± 50mVs, ± 500mVs

Range selection

Automatic or manual

Accuracy

± 1µVs after self calibration

Magentic Values

Magnetic Flux: Vs, Wb, Maxw
Magnetic Flux density: T, G
Magnetic Field strength: A/m, Oer
Magnetic Dipole moment: Am2, Vsm
Magnetic Potential: AMagnetic Flux: Vs, Wb, Maxw
Magnetic Flux density: T, G
Magnetic Field strength: A/m, Oer
Magnetic Dipole moment: Am2, Vsm
Magnetic Potential: A

Measurement

DC, AC uo to 2500Hz

Calibration

Internal voltage and time reference, self-calibration

Drift

Standby: 0 µVs/min after dedrift
Messzyklus: <± 1 µVs/min nach Dedrift Standby: 0 µVs/min after dedrift
Messzyklus: <± 1 µVs/min nach Dedrift

Input resistance

10kOhm ± 0.1%

Input terminals

4-polige LEMO-plug

Display

LCD 240*65 pixel, 127mm*34mm

Interfaces

Analogue Output

±10V, Continuous output of displayed readings
Peak (Min. Max. Diff.)±10V, Continuous output of displayed readings
Peak (Min. Max. Diff.)

Serial Interface

RS 232C, SCPI Code, 9600 Baud, 115kBaud via USB

SPS

8 inputs/outputs 24VDC, short circuit proof, potential free

Software function

 

Automatic drift compensation
Self calibration
8 different Go/No Go - Comparators
Storage of probe parameter in probe EEPROM

Helmholtz coils

Helmholtz coil 1D
Measurement of dipole moment of magnets

Helmholtz coil 3D
Measurement of dipole vector of magnets    

Measurement coils

Rotor Coils
Suited measurement coils for total flux of rotors

Point and Potential Measurement Coils
Measurement of local flux density and magnetic potential